Gravimetric measurement in redox polymer electrodes with the EQCM beyond the Sauerbrey limit

Gravimetric measurements during electrochemical redox switching of an Os-containing hydrogel film in contact with aqueous electrolyte have been obtained beyond the Sauerbrey rigid mass limit of the electrochemical quartz crystal microbalance (EQCM). For an acoustically non-rigid film the frequency-t...

Descripción completa

Guardado en:
Detalles Bibliográficos
Publicado: 1999
Materias:
Acceso en línea:https://bibliotecadigital.exactas.uba.ar/collection/paper/document/paper_13882481_v1_n5_p167_Etchenique
http://hdl.handle.net/20.500.12110/paper_13882481_v1_n5_p167_Etchenique
Aporte de:
Descripción
Sumario:Gravimetric measurements during electrochemical redox switching of an Os-containing hydrogel film in contact with aqueous electrolyte have been obtained beyond the Sauerbrey rigid mass limit of the electrochemical quartz crystal microbalance (EQCM). For an acoustically non-rigid film the frequency-to-mass Sauerbrey relationship is not valid and the mass has been obtained from the experimental electroacoustic impedance data collected simultaneously with electrochemical experiments interpreted using Martin's viscoelastic model.