Validation of a spectrophotometric method in the visible region for the quantification of kojic acid in raw materials and products

A spectrophotometric method in the visible region was developed and validated for the quantification of kojic acid (KA) in raw materials and products. According to the validation results, this method yields linearity in the range of 5 and 50 μg/mL of KA (r = 0.99998), selectivity, suitable sensiti...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Piantavini, Mário S., Stremel, Dile P., Trindade, Ângela C.L.B., Pontarolo, Roberto
Formato: Articulo Comunicacion
Lenguaje:Inglés
Publicado: 2010
Materias:
Acceso en línea:http://sedici.unlp.edu.ar/handle/10915/8092
http://www.latamjpharm.org/resumenes/29/8/LAJOP_29_8_2_9.pdf
Aporte de:
Descripción
Sumario:A spectrophotometric method in the visible region was developed and validated for the quantification of kojic acid (KA) in raw materials and products. According to the validation results, this method yields linearity in the range of 5 and 50 μg/mL of KA (r = 0.99998), selectivity, suitable sensitivity, precision (RSD 1.33 % and 1.21 % for intermediate precision and repeatability respectively), accuracy (recovery near 100 %) and robustness (varying pH, temperature and reading time) were measured and led to the validation of this method.These methods were tested and validated for various parameters according to ICH guidelines and USP.