Determination of sulphate for measuring magnesium sulphate in pharmaceuticals by flow analysis-fourier transforms infrared spectroscopy

The viability of employing flow analysis coupled to Fourier transform infrared spectroscopy (FA-FTIR) as a useful tool for quantitative measuring of magnesium sulphate in pharmaceuticals was explored, developed and validated. The method was based on mid-IR transmittance measurements of the peak-area...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Ovalles, Fernando, Gallignani, Máximo, Rondón, Rebeca, Brunetto, Maria R., Luna, Rafael
Formato: Articulo
Lenguaje:Inglés
Publicado: 2009
Materias:
Acceso en línea:http://sedici.unlp.edu.ar/handle/10915/7743
http://www.latamjpharm.org/trabajos/28/2/LAJOP_28_2_1_2_08C91ACD3F.pdf
Aporte de:
Descripción
Sumario:The viability of employing flow analysis coupled to Fourier transform infrared spectroscopy (FA-FTIR) as a useful tool for quantitative measuring of magnesium sulphate in pharmaceuticals was explored, developed and validated. The method was based on mid-IR transmittance measurements of the peak-area belonging to the sulphate band around 1110 cm–1 and the use of an external calibration curve. Dynamic range was established over a concentration range from 1 to 50 mg.ml–1 , with a limit of detection (3σ) of 0.26 mg.ml–1 The analytical frequency was 12 h–1 with a precision close to the unit (% RSD). The . analytical results obtained in commercial formulations by applying the proposed FA-FTIR method were in strong agreement with labelled values and those obtained by a reference titration method at 95% confidence level. Among various advantages offered by the proposed method over conventional ones, simple strategy and clean analytical chemistry must been highlighted.