A chopped front-end system with common-mode feedback for real time ECG applications
"A Front-End system composed of an instrumentation amplifier with common-mode feedback and a second-order sigma-delta analog to digital converter is proposed. Chopping techniques are used to remove offset and low frequency noise to guarantee a noise floor below the signal of interest. Measurem...
Guardado en:
| Autores principales: | , , , |
|---|---|
| Formato: | Ponencias en Congresos acceptedVersion |
| Lenguaje: | Inglés |
| Publicado: |
2019
|
| Materias: | |
| Acceso en línea: | http://ri.itba.edu.ar/handle/123456789/1715 |
| Aporte de: |
| Sumario: | "A Front-End system composed of an instrumentation amplifier with common-mode feedback and a
second-order sigma-delta analog to digital converter is proposed. Chopping techniques are used to remove offset and low frequency noise to guarantee a noise floor below the signal of interest. Measurements show that common-mode feedback adds 71 dB to the CMRR at 50 Hz and the input referred noise is 1.17 μVRMS over the frequency range of 0.1 Hz to 400 Hz." |
|---|