A high performance liquid chromatography method for determination of hydroxyzine hydrochloride in syrup
A simple isocratic method for quantification of hydroxyzine dihydrochloride by HPLC with UV detection at 232 nm in syrup has been developed and validated. Separation was achieved on a C18 column (250 x 4.6 mm, 5 μm) maintained at 90 ºC with 0.5 mol/L potassium dihydrogen phosphate buffer : acetonitr...
Guardado en:
| Autores principales: | , , , , , , , , |
|---|---|
| Formato: | Articulo |
| Lenguaje: | Inglés |
| Publicado: |
2011
|
| Materias: | |
| Acceso en línea: | http://sedici.unlp.edu.ar/handle/10915/8377 http://www.latamjpharm.org/resumenes/30/9/LAJOP_30_9_1_21.pdf |
| Aporte de: |
| Sumario: | A simple isocratic method for quantification of hydroxyzine dihydrochloride by HPLC with UV detection at 232 nm in syrup has been developed and validated. Separation was achieved on a C18 column (250 x 4.6 mm, 5 μm) maintained at 90 ºC with 0.5 mol/L potassium dihydrogen phosphate buffer : acetonitrile (1:1, v/v) as mobile phase at a flow rate of 2.0 mL/min. The method was proven to be linear over the range of 80–120 μg/mL, accurate (recovery = 97.8 %), precise (coefficient of variation = 1.52 % for sample) and robust. |
|---|